Voltage Fluctuation and Flicker Measurement (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Period during which relative voltage change exceeds the threshold level Tmax
The time during which the relative voltage change during a voltage fluctuation period
exceeds the threshold level.
Relationship between dc, dmax, and Tmax
For IEC 61000-4-15 Ed1.1
dmax
dc
Tmax
Threshold
level
Steady-state condition
Relative voltage change
For IEC 61000-4-15 Ed2.0
Threshold
level
Steady-state condition
Tmax
dmax
dmax +
dmax -
dc
Previous steady-state
condition
Relative voltage change
dmax +: The difference between the minimum value and the value in
the previous steady-state condition
dmax –: The difference between the maximum value and the value in
the previous steady-state condition
dmax: The larger of the absolute values of dmax + and dmax –