2-11
IM WT3001E-51EN
Built-in Printer (Option)
2
•  Printing or Not Printing (ON/OFF) the Header Information
10.
 Press the Information soft key to select ON or OFF.
     
•  Executing the Print Operation
11.
 Press PRINT to print the numeric data list.
•  Aborting the Print Operation
12.
 Press PRINT while the print operation is in progress to abort the operation.
Explanation
•  Header Information
  The following items are available as header information.
•  Measurement mode (Normal)
*1
• Date/Time
*1
•  Measurement range of each element
*2
•  PLL source
*2
  Turning ON/OFF Header Information Printing
•  ON: Prints the header information.
•  OFF: Does not print the header information.
*1 Even if Information is set to OFF, the measurement mode and date/time are 
always printed.
*2 Only installed elements are applicable for the selection.
•  Selecting the Numeric Data
•  Only installed elements and wiring units are applicable for the selection.
•  There are limitations on the numeric data that can be measured and printed in 
measurement modes
*3
 other than normal measurement mode. For details on the 
numeric data limitations, see appendix 11 in the User’s Manual IM WT3001E-01EN. 
[-------] (no data) is printed in places where there is no numeric data.
•  The harmonic order range is DC (0
th
) to the Max Order specified in section 7.5.
*3 Can be set on models with the advanced computation (/G6) option. Wide 
bandwidth harmonic measurement mode, IEC harmonic measurement mode, 
waveform computation mode, and FFT mode are available.
•  Print Format of the Numeric Data List
•  The selected numeric data items are printed.
•  For harmonic order data, the elements and measurement functions are printed 
horizontally, and the order list is printed vertically.
2.4  Printing Numeric Data Lists on the Built-in Printer