3-11
IM WT3001E-51EN
D/A Output and Remote Control (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Explanation
D/A Zoom
The D/A output can be magnified or reduced for each channel in manual range mode
by setting the D/A output range. For example, if a current varying in the range of
0.6 A to 0.8 A is measured at the 1-A range and the D/A output is set to fixed range
mode, the D/A output voltage varies between 3.0 V and 4.0 V. If you wish to magnify
this variation, you can use the D/A zoom function. In this example, set the D/A output
to manual range mode and set the maximum and minimum values to 0.6 and 0.8,
respectively. Then, the D/A output delivers –5 V and +5 V when the measured current
is 0.6 A and 0.8 A, respectively.
Selecting the D/A Output Range Mode
Select either mode from below.
• Manual
The displayed values of the measurement functions corresponding to the D/A
output of –5 V and +5 V can be set arbitrarily.
• Fixed
Outputs +5 V when the rated value of each measurement function is applied. For
details, see section 3.2.
Setting the Maximum and Minimum Values for Manual Range Mode
The maximum and minimum values can be set in the range of –9.999E+30 to
9.999E+30.
1.0 A
0.8 A
0.6 A
5.0 V
4.0 V
3.0 V
5.0 V
–5.0 V
Time
Time
Measured current
D/A output D/A output
Fixed range mode
Manual range mode
(DA zoom)
3.3 Setting the D/A Zoom