3-11
IM WT3001E-51EN
D/A Output and Remote Control (Option)
3
Explanation
•  D/A Zoom
  The D/A output can be magnified or reduced for each channel in manual range mode 
by setting the D/A output range. For example, if a current varying in the range of 
0.6 A to 0.8 A is measured at the 1-A range and the D/A output is set to fixed range 
mode, the D/A output voltage varies between 3.0 V and 4.0 V. If you wish to magnify 
this variation, you can use the D/A zoom function. In this example, set the D/A output 
to manual range mode and set the maximum and minimum values to 0.6 and 0.8, 
respectively. Then, the D/A output delivers –5 V and +5 V when the measured current 
is 0.6 A and 0.8 A, respectively.
•  Selecting the D/A Output Range Mode
  Select either mode from below.
• Manual
  The displayed values of the measurement functions corresponding to the D/A 
output of –5 V and +5 V can be set arbitrarily.
• Fixed
  Outputs +5 V when the rated value of each measurement function is applied. For 
details, see section 3.2.
•  Setting the Maximum and Minimum Values for Manual Range Mode
  The maximum and minimum values can be set in the range of –9.999E+30 to 
9.999E+30.
 
5.0 V
4.0 V
3.0 V
5.0 V
–5.0 V
Time
Measured current
D/A output D/A output
Fixed range mode
Manual range mode 
(DA zoom)
3.3  Setting the D/A Zoom