3-13
IM WT3001E-51EN
D/A Output and Remote Control (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Explanation
Hold, single measurement, integration start/stop/reset, and print output can be controlled
externally.
CAUTION
Do not apply voltage outside the range of 0 to 5 V to the remote control input pins.
Also, do not short the output pins or apply external voltage to them. If you do, the
instrument may malfunction.
French
ATTENTION
Ne pas appliquer de tension hors de la plage 0 à 5 V aux broches d’entrée de la
télécommande. Ne pas court-circuiter non plus les broches de sortie, ni y appliquer
de tension externe. Le cas échéant, un dysfonctionnement de l’instrument est
possible.
Remote Control Input/Output Circuit
0.01 μF 0.01 μF
+5 V
+5 V
10 kΩ
10 kΩ
100 Ω
Low level: 0 to 1 V
High level: 4 to 5 V
Input circuit Output circuit
Low level: 0 to 1.5 V (8 mA)
High level: 2.8 to 5 V (–8 mA)