11-22 Nightingale PPM3 User's Guide April 2019
Working With Menus
Parameter
Lower limit adjustment
Upper limit adjustment
HR/PR, RR,
ETCO
2
, RRc
NBPs/m/d,
P1s/m/d,
P2s/m/d
Changed to 80% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
Changed to 125% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
FICO
2
N/R
Changed to 125% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
IPI
Changed to 75% of the
current parameter value (or
the nearest allowable value
given the alarm limit settings
rules)
N/R
SpO
2
Changed to 95% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
Set to 100%
Temp
Changed to 95% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
Changed to 105% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)