12-10 Nightingale PPM3 User's Guide April 2019
Managing Alarms
PPM3 Battery Conditions:
Condition
Display
Value
Alarm
Grade
Message
Annunciation
Type
Battery okay
N/A
None
None
Persistent
Battery low
N/A
Low
Battery low
One time
HR Conditions:
Condition
Display
Value
Alarm
Grade
Message
Annunciation
Type
HR within limits
<Number>
None
None
Persistent
HR < LL
<Number>
Medium
HR < LL
Persistent
HR > UL
<Number>
Medium
HR > UL
Persistent
HR Asystole
ASY
High
HR asystole
Persistent
HR Ventricular Fibrillation
VF
High
HR ventricular fibrillation
Persistent
HR Lead-off (after
startup/standby)
<Blank>
None
None
Persistent
HR Lead-off (after leads applied)
<Blank>
Low
HR lead off
One Time
HR Artifact
---
Low
HR artifact
Persistent
RR Conditions:
Condition
Display
Value
Alarm
Grade
Message
Annunciation
Type
RR within limits
<NUM>
None
None
None
RR < LL
<NUM>
Medium
RR < LL
Persistent
RR > UL
<NUM>
Medium
RR > UL
Persistent
RR > 120 breaths/min
---
Low
RR out of range (too high)
Persistent
Lead-off (after start-up / standby)
<Blank>
None
None
Persistent
Lead-off (after leads applied)
<Blank>
Low
RR lead off
One Time
Resp Artifact
---
Low
RR artifact
Persistent

Table of Contents