12-14 Nightingale PPM3 User's Guide April 2019
Managing Alarms
Pressure signal out of range
(low)
<Number >
Medium
P1 out of range (low)
Persistent
No pulse rate due to static
pressure
---
Low
HR weak signal
Persistent
Unplugged (after start-
up/standby)
<Blank>
None
None
Persistent
Unplugged (after line connected)
<Blank>
Low
P1 unplugged
One Time
Zero required (when transducer
is initially connected)
<Blank>
None
P1 needs zeroing
Persistent
Zero required and 60 seconds
has expired.
---
Low
P1 needs zeroing
Persistent
Zero in progress
<Blank>
None
P1 zeroing
Persistent
Zero failed pulsatile waveform
N/A
Alert
P1 unable to zero - unstable
Alert
Zero failed out of range
N/A
Alert
P1 unable to zero - out of
range
Alert
Calibration required
<Blank>
None
P1 needs calibration
Persistent
Calibration required and 60
seconds has expired.
---
Low
P1 needs calibration
Persistent
Calibration in progress
<Blank>
None
P1 calibrating
Persistent
Calibration failed pulsatile
waveform
N/A
Alert
P1 unable to calibrate -
unstable
Alert
Calibration failed bad cal
resistor
---
Low
P1 problem detected
Persistent
Temperature Conditions:
Condition
Display
Value
Alarm
Grade
Message
Annunciation
Type
TEMP within limits
<NUM>
None
None
None
TEMP < LL
<NUM>
Medium
TEMP < LL
Persistent
TEMP > UL
<NUM>
Medium
TEMP > UL
Persistent
Temp > 50 degrees C
---
Low
TEMP out of range
Persistent
TEMP probe disconnected
---
Low
TEMP unplugged
One Time
TEMP bad calibration resistor
---
Low
TEMP problem detected
Persistent