24-12 Nightingale PPM3 User's Guide April 2019
Guidance and manufacturer’s declaration: electromagnetic immunity
The Nightingale PPM3 is intended for use in the electromagnetic environment specified below.
The user of the Nightingale PPM3 should assure that it is used in such an environment.
Electromagnetic environment /
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power
supply lines
1 kV for
input/output
lines
2 kV for
power
supply lines
1 kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
1 kV
differential
Mode
2 kV common
mode
1 kV
differential
Mode
2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
<5 % UT
(>95 % dip in
UT)
for 0.5 cycle
40 % UT
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
<5 % UT
(>95 % dip in
UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
Nightingale PPM3 requires continued
operation during power mains
interruptions, it is recommended that the
Nightingale PPM3 be powered from an
uninterruptible power supply or a fully
charged battery.