24-14 Nightingale PPM3 User's Guide April 2019
Technical Data
Guidance and manufacturer’s declaration: electromagnetic immunity
The Nightingale PPM3 is intended for use in the electromagnetic environment specified below.
The user of the Nightingale PPM3 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment /
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the Nightingale PPM3,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
, should be
less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)