April 2019 Nightingale PPM3 User's Guide 24-15
Technical Data
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Nightingale PPM3 is used exceeds the applicable RF
compliance level above, the Nightingale PPM3 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the PPM3 monitor.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.