April 2019 Nightingale PPM3 User's Guide 8-7
Masimo CO
2
Monitoring
WARNING (ISA) Never sterilize or immerse the ISA sidestream gas
analyzer in liquid.
WARNING (ISA) ISA sidestream gas analyzer is intended only as an
adjunct in patient assessment. It must be used in conjunction with
other assessments of clinical signs and symptoms.
WARNING (ISA) Measurements can be affected by mobile and
portable RF communications equipment. Make sure that the ISA
sidestream analyzer is used in the electromagnetic environment
specified in this manual.
WARNING (ISA) Replace the sampling line if the sampling line input
connector starts flashing red, or a "CO
2
Occluded Line” message
appears on the PPM3.
WARNING (ISA) No modification of this equipment is allowed
without authorization of the manufacturer. If this equipment is
modified, appropriate inspection and testing must be conducted to
ensure continued safe operation.
WARNING (ISA) ISA sidestream gas analyzers are not appropriate
for MRI environments.
WARNING (ISA) During MRI scanning, the ISA must be placed
outside the MRI suite.
WARNING (ISA) Use of high frequency electrosurgical equipment in
the vicinity of the ISA analyzer may product interference and cause
incorrect measurements.
WARNING (ISA) Do not apply negative pressure to remove
condensed water from the Nomoline family sampling line.
WARNING (ISA) Too strong positive or negative pressure in the
patient circuit might affect the sample flow.
WARNING (ISA) Strong scavenging suction pressure might affect
the sample flow.