WARNING (ISA) – Exhaust gases should be returned to the patient
circuit or to a scavenging system.
WARNING (ISA) – Due to the risk of patient cross-infection, always
use a bacteria filter on the exhaust port side if sampled gas is
intended to be re-breathed.
WARNING (ISA) – Do not place the ISA gas analyzer in any position
that might cause it to fall on the patient.
WARNING (ISA) – Do not re-use disposable single-patient use
Nomoline Family sampling lines due to the risk of cross
contamination.
WARNING (ISA) – Do not sterilize or immerse Nomoline Family
sampling lines in liquid.
WARNING (ISA) – Do not operate the ISA sidestream gas analyzer if
the enclosure is damaged.
WARNINGS (IRMA) – The IRMA probe is intended for use by
authorized and trained medical personnel only.
WARNINGS (IRMA) – The IRMA probe is intended only as an adjunct
in patient assessment. It must be used in conjunction with other
assessments of clinical signs and symptoms.
WARNINGS (IRMA) – The IRMA probe must not be used with
flammable anesthetic agents.
WARNINGS (IRMA) – Disposable IRMA airway adapters shall not be
reused. Reuse of the single use adapter can cause cross infection.
WARNINGS (IRMA) – Used airway adapters shall be disposed of in
accordance with local regulations for medical waste.
WARNINGS (IRMA) – Do not use the IRMA Adult/Pediatric airway
adapter with infants as the adapter adds 6 ml dead space to the
patient circuit.