Service Kit Installation Procedures
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 4–77
3 Remove the two Pan Head 4-40 x 1/4 screws on
the dovetail mounting bracket but do not remove
the two flat head screws. Disconnect the ribbon
cable on the Power Interface Module (PIM) board
by simultaneously applying pressure on the two
ejector latches.
Save these screws.
4 Flip the ventilator over and remove the front case
assembly by lifting it straight up away from the
ventilator module.
5 The O2 valve is located on the ventilator module.
6 The 6" corrugated tubing can be removed for
easier viewing and access to the O
2
Valve.
Step Procedure