Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–21
Troubleshooting — Oxygen System Calibrations
Remote Calibration System
Discrete Tests Test Purpose/ Definition Troubleshooting
O2 Kickstart
Calibration
Calibrates near zero flow target used for
fine flow control after opening valve.
(range: 50 - 900)
Typical failures are with the Oxygen Delivery
System. Gas flow is critical.
If gas flow is restricted this test may not successfully
calibrate.
Check the gas regulator to ensure tank is fully open.
Check set up
Replace Oxygen valve assembly kit.
O2 Kick Start
Check
Deliver 50 ml breath using 100% FiO2
O2 Flow
Calibration
OF_Offset
(range: 26053 - 37516)
OF_Gain
(range: 12727 - 34650)
Check set up
Gain Too Low – clean flow screens and re-install.
Check diffuser screen is present in assembly.
Gain Too High – check test setup and assembly for
leaks, then recalibrate. If issue persists, check
assembly to ensure O-rings were installed on either
side of the flow screen and diffuser screen is
properly installed.
Replace the SPM/Vent assembly kit.
Flow Calibration
Check
Compares selected flow values to published
specifications (incoming), or tighter
(outgoing) to allow for drift.
Ventilator Alarms
Title Service
Code
Alarm Purpose Troubleshooting
Self-Check: Transducers/ Calibration Files
Self Check Failure /
Self Check Fault
1010
2010
O2 Valve Stuck Open
Connect O2, listen for flow.
If no flow can be heard, recalibrate device.
Replace O2 Valve
Self Check Failure /
Self Check Fault
1011
2011
Control fault to indicate
Insufficient O2 Flow rate.
Recalibrate device
Check manifold for leaks
Replace O2 Valve
Check Flow Screen Assembly, clean and replace.
Self Check Failure 1012 PGA or ADC faults detected
for O2 System.
Recalibrate Device
Replace SPM
Low 02 Supply Failure /
Low 02 Supply Fault
1020
2020
Oxygen Tank Pressure low
(less than 35 psig)
Reduce O2 Supply Pressure
Power Cycle Device
Check Oxygen Regulator
Replace SPM if Supply Pressure is verified to be
below 87 psig.
High 02 Supply Failure 1041
2041
Oxygen tank pressure is
excessive (greater than 87
psig)
Reduce O2 Supply Pressure
Power Cycle Device
Check Oxygen Regulator
Replace SPM if Supply Pressure is verified to be
below 87 psig.
Self Check Failure 1174 Flow transducers are not
behaving as expected.
Change SPM board
Self Check Failure 1176 Checks for a corrupted
calibration file.
Calibration required.
Observations
Observation Resolution