Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–29
Service Codes - Version 5
A list of all ventilator Service Codes for Version 5follows and is divided into three categories:
• High Priority (1XXX)
• Medium Priority (2XXX)
• Low Priority (3XXX)
Service Codes in each category are provided in numerical order.
NOTE:
Before ordering a service kit, ensure that all troubleshooting steps listed earlier in this chapter have been followed.
CODES NAME
Description /Possible
Cause
Actions
1001
2001
3001
Compressor Flow
Path / Failure
Description:
Compressor does NOT
respond to control
signals
Possible Cause:
• Blocked output
• Tubing kinked
• Connector loose
• Diffuser or flow
screen issue
1. Does not present at power up
• Use Conditions (Possible settings conflict or the output
was blocked during operation):
• Download forensic memory using RCS consult technical
support; possibly the output was blocked.
2. Presents at power up
• Check tubing -> tubing kinked or disconnected-> Correct
tubing
• Check compressor cable is properly seated -> dislodged
connector -> re-seat connector
• Check diffuser screen, check flow screen ->collapsed
diffuser screen or wet flow screen - replace Flow Screen
Kit
1002
2002
3003
Internal
communication
Compressor
(Failure/Faults)
Description:
Compressor flow signal
- High, Low, or Stuck
Possible Cause:
• Tubing kinked or
disconnected
• Faulty compressor
flow transducer
• Check tubing -> tubing kinked or disconnected-> Correct
tubing
• Check Flow transducer -> Bad flow transducer ->
Replace Smart Pneumatic Module (SPM)/ Vent
Assembly Kit
1003 Sensor/
Transducers/Cal
file Failures
Description:
Fails power-on self
check breath
Possible Cause:
• Calibration error
• Foreign substances
in screen
• Faulty compressor
flow transducer
• calibration Error -> Calibrate
• check diffuser screen check flow screen ->collapsed
diffuser screen, foreign substances contamination or
wet flow screen - replace screen
• Bad flow transducer -> Replace SPM Module