Service Kit Installation Procedures
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 4–81
Flow Screen Service Kit Installation Instructions
Kit Number (ALL MODELS)
712-0731-41
SCAN OR CLICK FOR VIDEO
Compressor Flow Screen
O2 Flow Screen
CAUTION / WARNING!
Replacement of the Flow Screen requires the handling of Oxygen Clean Parts. When handling any metal
components within the ventilator which are responsible for transporting oxygen, take the following precautions:
• Never touch the components with your bare hands as oils and residue may contaminate the
component.
• Keep all Oxygen Clean parts in their protective packaging prior to installation.
• Do not keep sources of contaminate on the workbench such as soldering material.
• If an oxygen clean part becomes contaminated, replace the part and do not attempt to clean it.