EasyManua.ls Logo

ZOLL EMV+ - Part III - Service Codes; Retrieving Service Codes Version 5 Ventilators only

ZOLL EMV+
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ventilator & Calibration Troubleshooting
6–28 www.zoll.com 906-0731-04-01 Rev. C
Part III — Service Codes
This section provides a list of Version 5 and Version 4 service codes of service related alarms, their associated alarm
description and their resolution for restoring proper ventilator operation.
A service Code is a four-digit number associated with each ventilator alarm.
The code appears at the bottom of the Alarm Message Center (AMC). In Release 5.x ventilators the Service Code is
also listed on the Alarm History.
Retrieving Service Codes
Version 5 Ventilators only
Retrieving Service Codes (Alarm History) is accomplished by the Read Forensic Memory sequence on RCS software.
Running this sequence will read the alarm history data from the ventilator and parse that data into five files. The five
files are:
• Alarms
• Breaths
• Conditions
• Settings
• Summary (AllData - RCS 8.X)
The files are written in CSV (comma separated values) format and can be viewed through any text editor or
spreadsheet application.
There will be sub-folders named with the embedded serial number of the vent (EMV Model + EMV SN) ex.
VCDB123456.
Each time the Read Forensic Memory is run, five additional files will be written to the directory location. The file
names will include the current date and time to prevent overwriting previously read Forensic Memory files.