Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–27
Ventilator_Memory_Logs
Documents\ZOLL\Ventilator_Me
mory_Logs\”EMV Model & SN”\
This folder will contain several files and is created when the forensic
memory is read from the ventilator. This folder is only present after the read
forensic memory is executed. The files document device settings and
conditions as well as all alarms active on the device at the time of each
entry. Device logging is cyclic, meaning once the forensic memory is full on
the device, it will continue to log overwriting the oldest data first. The log
does not discriminate against alarms generated by use of RCS vs. those
generated on the device while it is in use and all time is formatted in UTC
time. The log may contain alarms that were not present on the display when
used with RCS, which is expected behavior. When communicating with the
SPM, communication is broken with the EMV. The opposite is also true.
Therefore, it is common and expected to have communication alarms
present in forensic memory after conducting tests with RCS.