906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 5–1
This chapter provides information on the following:
About Electrical Safety Testing
Warnings
Electrical Safety Testing Procedure
Touch Point Diagrams
About Electrical Safety Testing
Electrical safety testing is non-destructive testing that confirms the suitability of electrical insulation. Electrical
safety testing is helpful for discovering damaged insulation, stray wire strands or damaged braided shielding.
Refer to the manufacturer’s instructions or supplied specifications for the leakage tester you use. Read the
electrical safety testing device user and operator’s manuals to become familiar with the device (safety
warnings, cautions, safety markings) and to confirm that the testing area (station) where the electrical safety
testing device operates conforms to all guidelines, safety warnings, and cautions.
Perform these tests at the line-power voltage and frequency used in your installation. The 731 Series Ventilator
NOTE:
Due to the wide variety of electrical safety test equipment on the market, it is not possible to provide detailed step
by step instructions for this procedure.