Service Kit Installation Procedures
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 4–95
Cleaning Instructions - Mesh Screen
• You may wash the Flow Screen with a Liquinox or equivalent solution.
• You may use an ultrasonic bath with Liquinox or equivalent solution. See information here: https://
alconox.com/liquinox/
• Allow the Flow Screens to air dry before re-inserting them into the Manifold. DO NOT use compressed
air from an oil-based compressor.
• Inspect the screen against a light to verify cleanliness before re-inserting them into the manifold