Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–23
Troubleshooting — Test Station
Remote Calibration System
Discrete Tests Test Purpose Troubleshooting
TSI Cal Check
Confirm the TSI calibration,
Used within temperature and altitude
range.
1) TSI Barometric Pressure: 69kPa <
reading < 102kPa (~10,000 ft. to -200 ft.),
2) TSI Temperature: 16C < reading < 32C,
3) TSI Calibration: current date does not
exceed due date.
Return to main menu, select “Get TSI Flow Module”
information button and confirm operating conditions
Swap-out to new module;
Use notes to justify usage when out of spec.
Observations
Observation Resolution
Prior to servicing the device, ensure equipment and
technician safety.
Use proper ESD precautions at all test stations when
opening the device.
Follow proper safety precautions as noted in the
Service Manual
The tool is not compliant to FDA Part 11
Print hard copy of each record, sign and maintain in
accordance to Quality System.
PC Setup
If this is a new installation and if the drivers did not
register properly then the user will be unable to
read/write to the ventilator’s firmware.
If connected to the Internet the Atmel driver may be
replaced by windows automatically to a non-
compatible open-sourced driver (BOSSA). Fix:
Reinstall the TCE software.
With unit plugged in, verify “Load VCP” is unchecked
for both USB Serial Converter properties advanced
tab.
Note: To uncheck the “Load VCP” option, follow the below steps:
Navigate to: device manager > universal serial bus controllers > USB serial converter
Right click and choose properties, and select the advanced tab.
Verify that Load VCP is unchecked.
Repeat these steps for the 2nd USB serial converter.
TSI flow modules need to be calibrated on a yearly basis
Contact TSI and calibrate module on an annual basis.
(See RCS manual)
The TSI may not maintain calibration for the year.
Establish a monthly flow check.
Properly store and protect the TSI module.
TSI Cables
Inspect the TSI data cable to avoid data interruptions.
TSI Data Cable
O2 Cable
O2 Cell
Calibrate when prompted
Faulty O2 Regulators
The O2 regulators need to be properly used to ensure proper flow.
Oxygen Regulators may be damaged due to improper
use.
Always close the tank and relieve pressure from both
sides of the regulator when not in use.
Always open the tank slowly with the regulator in the
completely closed position (0 PSI). Turn the regulator
handle fully counter clockwise.