Ventilator & Calibration Troubleshooting
6–24 www.zoll.com 906-0731-04-01 Rev. C
O2 Supply
The oxygen supply used with RCS is critical to the
ability to calibrate and test devices.
• It is recommended to perform frequent checks.
• Confirm O2 regulated supply (see Diagnostics Section
below)
• The supplied O2 hoses use DISS 1240 fittings and O
ring seal.
• It is recommended to hand tighten the O2 Fittings,
they do not need to be tightened with a wrench and
may be damaged if overtightened.
• Debris may become lodged in the tubing from
damage to the O-rings.
• Inspect hoses to ensure O-rings is in place and not
damaged
Hidden Pop Up.
Users waiting for Prompt.
• Prompt is hidden by application.
• Minimize Window to see Pop UP.