Ventilator & Calibration Troubleshooting
6–10 www.zoll.com 906-0731-04-01 Rev. C
Status Window
The window will indicate a device has been connected:
Open Port
Close Port
The window will indicate when power is applied:
• Use the status window to troubleshoot
communication problems.
• Power Cycle needed
• Check USB – PC
• Check USB – Cable
• Check Hardware COMM
• Check TSI COMM (See Test Station)
• Retry to check if communication is established.
• The log files store the status window content.
When initiating the testing, there is a persistent message: FAILED TO
WRITE TESTLOAD.BOOTLOADER.BIN
The RCS does cannot read the device Serial Numbers, and the test load
cannot be installed on the device.
• If multiple failures occur at one station:
• Check USB Cable.
• OR
• Check USB-PC
• Check for Device Hardware COMM