Ventilator & Calibration Troubleshooting
6–10 www.zoll.com 906-0731-04-01 Rev. C
Status Window
The window will indicate a device has been connected:
Open Port
Close Port
The window will indicate when power is applied:
Use the status window to troubleshoot
communication problems.
Power Cycle needed
Check USB – PC
Check USB – Cable
Check Hardware COMM
Check TSI COMM (See Test Station)
Retry to check if communication is established.
The log files store the status window content.
When initiating the testing, there is a persistent message: FAILED TO
WRITE TESTLOAD.BOOTLOADER.BIN
The RCS does cannot read the device Serial Numbers, and the test load
cannot be installed on the device.
If multiple failures occur at one station:
Check USB Cable.
OR
Check USB-PC
Check for Device Hardware COMM