Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–15
Exhalation Backup and
Autocal Valve
Control Valves are exercised to ensure they
open and close as expected.
For both 4 cmH2O and 25 cmH2O pressure
targets, normalize waveforms and verify the
following:
1) Autocal rise settling time < 30ms
2) Autocal fall settling time < 30ms (<150ms
RCS 8)
3) Exhalation valve rise delay < 45ms
4) Exhalation valve fall delay < 45ms
5) Backup valve rise delay < 45ms
6) Backup valve fall delay < 45ms
7) Pressure difference between exhalation
valve drive port and airway pressure < 2.0
cmH2O.
Check setup for leaks or assembly error check lung.
Change the SPM Assembly.
Breath Performance
Testing
Checks ventilator tidal volume and selected
parametric values.
Target Settings:
VT=700; FIO2=21; BPM=20; Itime=1000;
PEEP=0
VT=100; FIO2=21; BPM=30; Itime=667.67;
PEEP=5
VT=700; FIO2=60; BPM=20; Itime=1000;
PEEP=0
VT=100; FIO2=60; BPM=30; Itime=667.67;
PEEP=5
VT=700; FIO2=100; BPM=20; Itime=1000;
PEEP=0
VT=100; FIO2=100; BPM=30; Itime=667.67;
PEEP=5
Tolerances:
VT: ± (5ml + 8% of setting)
PEEP: ± (2cmH2O + 7.5% of setting)
PIP: ± (2 cm H2O + 7.5% of setting)
BPM: ±1
FIO2: ± (3% + 8% of setting)
I Time: ± (100ms)
NOTE: For mixed flows, only FIO2 is checked
due to calibration issue with TSI
Check Set Up
Remove Resistor from compressor intake.
Remove High Pressure 02 Orifice / Resistor left on
the device.
Check test lung for proper inflation and uniformity
Check O2 cell for leaks, it's a two piece assembly
which may unscrew or leak
Check 22ml bacterial filter for cracks, leaks
ensure that the circuit is laying flat on the work
bench, with the exhalation valve in the upright
position
flush the o2 circuit before any retry, recal the O2
sensor.
Replace the O2 sensor if FIO2 issues persist.
If the vent can't capture PEEP, check the Anti-
Asphyxia valve for damage or debris
Recalibrate the ventilator