Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–35
3143 Self Check Fault Description:
Temperature Sensor
Fault displayed
Possible Cause:
1 of the 4 PGA's
temperature
measurements
beyond threshold
SPM Module is
faulty
Check to see if the error clears on power cycle ->Power
Cycle device if the error persists -> Replace SPM Module
1172 Self Check Failure Description:
5V Self Check Failure
Possible Cause:
PIM Board
malfunction
Any board in the
front panel
assembly or SPM
module
Check to see if the problem is within the SPM assembly
by disconnecting the front panel and powering up the
SPM, look for a blinking LED on the SPM PCB if no
blinking LED-> replace SPM Module
Check to see if the problem is within the SPM assembly
by disconnecting the front panel and powering up the
SPM, look for a blinking LED on the SPM PCB if the SPM
LED is blinking, then the problem resides within the
front panel assembly-> replace front panel assembly
1173 Self Check Failure Description:
Communication Failure
Possible Cause:
PIM to CPU ribbon
damage
CPU Module faulty
Check for damage/seating of the PIM to CPU cable -> if
the cable is damaged -> Replace PIM to CPU cable
Replace the PIM to CPU cable and the problem persists -
. CPU Module faulty-> Replace Front Panel Assembly
1174 Self Check Failure Description:
Unable to zero/null
transducer offset on
one or more
transducers
Possible Cause:
kinked tubing
transducer offset
drift (new offset too
far from calibrated
offset)
board
contamination /
bad transducer
Check to see if the transducer(s) tubing is kinked ->
tubing kinked -> Correct tubing
Check to see if the calibration has drifted by
recalibrating-> if the problem has resolved the issue was
transducer drift if the problem persists -> Replace the
SPM Module
Check for contamination on the SPM Module or bad
transducer -> SPM board contaminated/ Transducer bad
-> Replace SPM Module
1175 Self Check Failure Description:
Loss of communication
with any I2C device for
4 consecutive attempts
(on the SPM)
Possible Cause:
Stuck I2C bus
Check to see if the problem will clear on power cycle by
power cycling the unit-> if the problem has resolved the
I2C was stuck, if the problem persists -> Replace SPM
Module