Ventilator & Calibration Troubleshooting
6–48 www.zoll.com 906-0731-04-01 Rev. C
2090
PEEP Leak
Alarm triggers when the airway pressure drops below the PEEP setting by 2
cm H
2
O during the expiratory phase of the breath. This can be caused by a
leak in the breathing circuit, exhalation valve or patient airway.
Check internal/external
tubing connections
2095
Insufficient Flow
Alarm triggers when the pressure target is not reached during the
Inspiratory period during pressure targeted ventilation. Typically this can
occur when the Rise Time is set too low for the patient and their respiratory
mechanics.
Check internal/external
tubing connections
2100
Patient Disconnect
Alarm triggers when the airway pressure fails to exceed the PEEP setting by
~7 cm H
2
O.
Check internal/external
tubing connections
2110
Patient Detected
An alarm triggers when you connect the patient to the ventilator while the
Start Menu is still active. To resolve the alarm, you must select a mode of
ventilation and configure the device appropriately for the patient. In
addition, you should perform the Operational Test procedure before
reconnecting the patient to the device.
Mitigation/Info: Backup Ventilation Started, Set Mode (AC,
SIMV, CPAP, BL), Configure Other Settings, **Manually
Ventilate Patient and Restart**
Select a mode of ventilation.
Configure appropriate
settings
2170
Spont. Breath PIP High
Alarm triggers when the airway pressure, PIP, exceeds the High PIP Limit
Setting during 2 consecutive spontaneous breaths.
Check accuracy of Paw
Smart Pneumatic Module/
Vent Assembly Service Kit
2171
Spont. Breath PIP Low
Alarm triggers when the airway pressure, PIP, exceeds the Low PIP Limit
Setting during 2 consecutive spontaneous breaths.
Check accuracy of Paw
Smart Pneumatic Module/
Vent Assembly Service Kit
2172
Spont. Breath Vt High
Alarm triggers when the high VT Limit is exceeded during 2 consecutive
spontaneous breaths.
Smart Pneumatic Module/
Vent Assembly Service Kit
2173
Spont. Breath Tt Low
Alarm triggers when the Low VT Limit Setting is not achieved during 2
consecutive spontaneous breaths.
Smart Pneumatic Module/
Vent Assembly Service Kit
2300
Self Check Fault
Alarm triggers when the pulse oximeter module fails while in use. When the
alarm is active “-- --” appears in the HR and SpO
2
windows. Pressing the
Mute/Cancel button silences the audible alarm for 30 seconds.
CPU/UIM and SPO Stack
Service Kit
2301
Self Check Fault
Alarm triggers when the communication between the pulse oximeter
module and device fails.
CPU/UIM and SPO Stack
Service Kit
Service Code Associated Alarm Name/Description Service Kit/Resolution