Ventilator & Calibration Troubleshooting
6–50 www.zoll.com 906-0731-04-01 Rev. C
2430
Nearly Drained Battery
Alarm triggers when the device detects that there is ≤ 5 minutes of battery
operation remaining and external power is not connected.
4 Year Maintenance Kit
2450
Battery Discharge Fault
Alarm triggers when the battery temperature reaches 70 °C (158 °F) which is
5 °C from its maximum operating temperature using the internal battery
and external power is not connected. When the battery temperature
reaches 75 °C (167 °F) the battery will shut down to prevent failure and the
device will sound a high priority alarm and shutdown.
4 Year Maintenance Kit
2455
Battery Fault
Alarm triggers when the device is not able to communicate with the
internal battery. When this occurs the device does not know the current
charge of the battery and operation could stop at any time.
4 Year Maintenance Kit
Service Code Associated Alarm Name/Description Service Kit/Resolution