Maintenance Schedule & Procedures
3–10 www.zoll.com 906-0731-04-01 Rev. C
Table 1 - Test Sequences RCS 7.X
<------ RCS RELEASE 7 ------>
Discrete Test
Incoming System
Testing
Complete
Calibration
Outgoing
System Testing
Comm RAM and Bootloader Check ● ●
Power Switch Check ● ●
EMV Self Check ● ●
LCD Contrast Set ● ●
LCD Visual Inspection ● ●
Button Check ● ●
Total Power Failure Alarm Check ● ●
LED Check ● ●
Encoder Check ● ●
Buzzer Check ● ●
Li-Ion Battery Check ● ●
Pulse Oximeter SpO2 and Heart Rate ● ●
SPM Communication ●
SPM Self Check ● ● ●
SPM Blink ● ●
TSI Calibration Check ● ● ●
Calibration Check Barometric Pressure ● ●
Compressor Calibration ●
Incoming Calibration Check Compressor System ●
Calibration Check Compressor System ●
Airway Pressure Calibration ●
Incoming Calibration Check Airway Pressure ●
Calibration Check Airway Pressure ●
O2 Kickstart Calibration ●
Calibration Check O2 Kickstart ● ● ●
O2 Flow Calibration ●
Incoming Calibration Check O2 System ●
Calibration Check O2 System ●
O2 Leak Filter and Pressure Cal. Check ● ●
Manifold Leak Check ● ●
Exhalation Backup and Autocal Valve ● ●
Compressor Bypass Protection ● ●
Breath Performance Check ●
Date and Time Set ●
Date and Time Check ●
Download Calibration Tables ●
Verify UUT and Reference Device ● ● ●
Conditional PM Due Reset ●