8
2.4
Environment Requirements
Operating environment, storage environment and running environment:
Table 2-3 Overall environment requirements
Operating Environment
Requirements
Storage Environment
Requirements
Running Environment
Requirements
Ambient
Temperature
15℃~35
-10℃~40
10℃~40
Relative
Humidity
20%85%
10%90%
10%90%
Atmospheric
Pressure
70kPa106kPa
50kPa106kPa
70kPa106kPa
2.5
Product Specifications
2.5.1 Sample mode
Two sample modes are provided: WB (whole blood) mode and PD (pre-diluted) mode.
2.5.2 Throughput
The throughput of Z3 WB/PD is 70 samples/hour.
2.6 Testing Parameters
The analyzer provides quantified results for 21 report parameters (WBC, RBC, PLT, HGB, etc.)
and 3 histograms (WBC, RBC, and PLT). See the table below for details.
Table 2-4 Parameters
Name
Abbreviation
White Blood Cell count
WBC
Lymphocyte number
Lymph#
Mid-sized Cell number
Mid#
Granulocyte number
Gran#
Lymphocyte percentage
Lymph%
Mid-sized Cell percentage
Mid%
Granulocyte percentage
Gran%
Red Blood Cell count
RBC
Hemoglobin concentration
HGB
Mean Corpuscular Volume
MCV
Mean Corpuscular Hemoglobin
MCH

Table of Contents