Software and Interface
27
4.4
Gain Setup
You can set up the gain for HGB on the "Gain Setup" screen. Gains for other parameters are
obtained by gain calibration and cannot be edited.
Figure 4-4 Gain Setup
Adjust the HGB gain by clicking + or button to change HGB gain value to get HGB blank value around
4.5±0.1V, then save HGB gain.
As the gain settings affect the validity of analysis results, be careful when you adjust them.
NOTE