Reconnect communication cable or
replace PCBA board.
AD out of range [2.44-2.55]V
The time indicated by the system
clock is earlier than 2000-01-01
1. Check whether the button cell of
the main control board is installed;
2. Replace with a new button cell, and
set up the date and time in the setup
screen. Save the settings, shut down
the analyzer and then restart;
3. If the error still exists after
restarting the analyzer, replace the
main control board.
Fail To Exit Sleeping
Mode
System time is later than expire time
Check reagent expire time, replace
and prime diluent.
System time is later than expire time
Check reagent expire time, replace
and prime Lyse.
Floater status:
1. Check once in startup, and the
status is full;
2. Check for 3 times when the
analyzer is idled, and the results
all show that the status is full.
Check the floater, connecting cable.
The reagent detection reports no
diluent
The reagent detection reports no
diluent
Digital board 56V out of range
[47.0,63.0] V, constant current:
51.5V ~ 61.5V
Digital board +12V out of range
[11.0,13.0]V, +12V: 10.5V ~ 13.5V
Digital board -12V out of range [-
14.0,-9.0]V, -12V:-13.5V ~ -10.5V
Blank reading is out of range,
WBC≤0.2 *10^9 / L,
RBC≤0.02*10^12/ L,HGB≤1 g / L,
HCT≤0.5 %,PLT≤5*10^9 / L
HGB Blank Voltage
Abnormal
1, HGB voltage out of range [3.85,
4.85]V
2, HGB higher than 4.85V
Adjust HGB gain or maintenance
WBC chamber.
Vacuum Pressure is not in range