10
control, dilution ratio: 1:10) which is within the range specified in Table 2-8, test it consecutively for
3 times, and the test results are j1, j2, and j3. Calculate the carryover according to the following
equation, and the result shall meet the requirements in Table 2-7.
Table 2-7 Carryover Requirements
Parameter
Carryover
WBC
0.5
RBC
0.5
HGB
≤0.5%
PLT
≤1.0%
Table 2-8 Sample Concentration Range of Carryover Test
Parameter
Unit
High concentration range
Low concentration range
WBC
×10
9
/L
> 90.00
< 3.00
RBC
×10
12
/L
> 6.20
< 1.50
HGB
g/L
> 220
< 50
PLT
×10
9
/L
> 900
< 30
2.7.3 Repeatability
Test a sample which meets repeatability requirement on the analyzer consecutively for 10 times,
calculate the CV(%) and absolute deviation (d) of each parameter, and the results shall meet the
requirements in the following table.
In the equation:
s
----standard deviation of sample test results;
x
----mean value of sample test results;
xi
----actual test result of the sample;
d
----absolute deviation of the sample test results.