Aperture signal does not stable
Aperture Voltage
Abnormal
Aperture voltage is lower than 14V
Maintenance chamber or replace
main control board.
Impedance Signal
Interference
1.Signal interference
2.PLT noise signal more than 10%
Aperture signal does not stable
Aperture Voltage
Abnormal
Aperture voltage is lower than 14V
Maintenance chamber or replace
main control board.
Impedance Signal
Interference
1.Signal interference
2.PLT noise signal more than 10%
Thermal printer is
out of paper
Wait for a while and print again.
Thermal printer pole is not installed
well
Close thermal printer cover
Restart instrument, replace main
control board or motor drive board.
Fluidics Has Not
Initialized
Restart instrument, replace main
control board or motor drive board.
Restart instrument, replace main
control board or motor drive board.