75
Clog
0x01000901
Aperture signal does not stable
Unclog and test blank
Aperture Voltage
Abnormal
0x01000902
Aperture voltage is lower than 14V
Maintenance chamber or replace
main control board.
Impedance Signal
Interference
0x01000903
1.Signal interference
2.PLT noise signal more than 10%
/
Clog
0x01000904
Aperture signal does not stable
Unclog and test blank
Aperture Voltage
Abnormal
0x01000905
Aperture voltage is lower than 14V
Maintenance chamber or replace
main control board.
Impedance Signal
Interference
0x01000906
1.Signal interference
2.PLT noise signal more than 10%
/
Thermal printer is
out of paper
0x08000303
Lack of printer paper.
Replace printer paper
Thermal Printer
Error
0x08000304
Thermal Printer too hot
Wait for a while and print again.
Thermal printer
open
0x08000305
Thermal printer pole is not installed
well
Close thermal printer cover
Startup fail
0x09000001
Restart instrument, replace main
control board or motor drive board.
Fluidics Has Not
Initialized
0x09000002
Restart instrument, replace main
control board or motor drive board.
Cell counting
Abnormal
0x09000003
Restart instrument, replace main
control board or motor drive board.