42/24-10 EN Rev. 9 Chapter 7: Gas Analyzer Configuration 7-C-7
Adding a System Module
If no system module is configured or an added system module is not yet config-
ured in a gas analyzer, the NEW softkey will appear on the screen. Press this
softkey to go directly to the Setup System Modules menu.
The procedure for adding an analyzer module and an external I/O device differs
from the procedure for adding an I/O module (see the following instructions).
Step Action
1 Select the Set Up System Modules menu item.
The list of the system modules existing in the system is displayed.
2 Press the NEW softkey.
3 Enter the 14-digit serial number for the new system module.
4 In the list the new system module will be shown with a New status.
Adding a New
Analyzer Module or a
New I/O Device
5 Save the configuration change by pressing ENTER or discard it by
pressing Back.
Step Action
1 Select the Set Up System Modules menu item.
The list of the system modules existing in the system is displayed.
2 Select the I/O module which has been added and automatically
recognized by the gas analyzer and press the NEW softkey.
3 When adding a digital I/O module: Press the FB APPL. softkey and
select a function block application.
4 In the list the new system module will be shown with a New status.
Adding a New
I/O Module
5 Save the configuration change by pressing ENTER or discard it by
pressing Back.
When retrofitting a Profibus module, it must always be installed in the lowest slot
X20/ X21.