42/24-10 EN Rev. 9  Chapter 7: Gas Analyzer Configuration  7-A-7 
Limit Value Monitor Initialization 
 
Menu Path  MENU → Configure → Component specific → Limit Values → 
Select Limit Monitor → … 
 
Selection  All available limit value monitors are shown. 
 
Parameter Explanation  Action 
Direction  < =  Alarm on underflow threshold value 
> =  Alarm on exceeding threshold value 
Select 
Threshold in physical units  Set 
Procedure 
Hysteresis  in physical units  Set 
 
Standard 
Configuration 
As a rule, limit value monitoring for those components to be measured by the gas 
analyzer is factory-set. This requires that there be enough digital outputs on the 
I/O modules to handle the number of sample components. 
 
Note  Limit value monitors are factory-set or user-configured Limit Monitor function 
blocks. Technical Bulletin “AO2000 Function Blocks – Descriptions and Configu-
ration” (publication number 30/24-200 EN) contains complete information on the 
function block.