EasyManua.ls Logo

ABB Advance Optima magnos206 - Limas11, Uras26: Relinearization

ABB Advance Optima magnos206
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-8 Chapter 9: Inspection and Maintenance 42/24-10 EN Rev. 9
Limas11, Uras26: Relinearization
When should
relinearization be
performed?
Sample component relinearization should be performed
If the linearity deviation exceeds the permissible 1% of span
If the start of a suppressed measurement range is to be calibrated
After a component (lamp/ emitter, sample cell, calibration unit/cell, detector) is
installed in or removed from a beam path.
We recommend to perform the relinearization of a sample component after
measurement range limits have been changed (see “Limas11, Uras26: Notes for
Changing Measurement Range Limits” section, page 7-A-3).
Test Gases Depending on the number and type of measurement ranges, test gases with
varying concentrations are required for relinearization.
Number and Type of Meas. Ranges Test Gas Concentration
1 Measurement Range Approx. 40–60% of the measurement
range end value (“center point gas”)
2 Measurement Ranges End value of smaller measurement range
2 Measurement Ranges,
1 of which is suppressed
Start value of suppressed measurement
range
Menu Path MENU
Maintenance/Test Analyzer spec. adjustm.
Relinearization
Step Action
Relinearization is performed separately for each sample
component.
1 Perform the basic calibration for the zero and span of the sample
component to be relinearized (see page 9-30).
2 Select the Relinearization menu item.
3 Select the Sample component.
4 Turn on the test gas supply.
5 Enter the test gas concentration set point value.
6 Wait until the measurement value reading stabilizes and activate the
alignment procedure.
Procedure
7 Repeat steps 3-6 for all sample components.

Table of Contents

Related product manuals