42/24-10 EN Rev. 9 Chapter 9: Inspection and Maintenance 9-7
Limas11, Uras26: Measurement of Calibration Cells
Definition The measurement of a calibration cell in the Limas11 and Uras26 analyzer modules
means:
Determining what calibration cell “deflection” is equivalent to the test gas
calibration reading. This “deflection” is stored as the calibration cell “set value”.
When should
calibration cells be
measured?
We recommend measuring the calibration cells once a year.
We recommend measuring the calibration cells
after end-point calibration of a sample component with test gas or
after any change in measurement range limits (see “Limas11, Uras26: Notes for
Changing Measurement Range Limits” section, page 7-A-3) or
after a relinearization (see “Limas11, Uras26: Relinearization” section, page 9-8).
Before measuring the
calibration cells
Prior to measuring the calibration cells, the zero and end points of the applicable
sample components must be calibrated with test gases.
Test Gas During calibration cell measurement the zero gas supply must be turned on.
Menu Path MENU
Maintenance/Test Analyzer spec. adjustm.
Measure cal. cell