8-A-6 Chapter 8: Gas Analyzer Calibration 42/24-10 EN Rev. 9
Test Gas Supply Control for Automatic Calibration
Test Gas Supply The test gas supply for automatic calibration can be started by means of the gas
module’s integral solenoid valves or via external solenoid valves.
Depending on the gas supply circuit and the number of analyzer modules installed
there are several test gas layout possibilities (see Figures 8-A-2 to 8-A-6):
1 Analyzer Module,
Gas Module Installed
with 1 Solenoid
To calibrate analyzer modules with simplified calibration procedures:
Caldos27 with single-point calibration
Magnos206 with single-point calibration
Limas11 with calibration cells
Uras26 with calibration cells
Oxygen Sensor
Figure 8-A-2
Sample gas
Tes t gas
Analyzer
module
Electronics
module
System bus
1 Analyzer Module,
Gas Module Installed
with 3 Solenoids
To calibrate all analyzer modules that can be used with the gas module.
Figure 8-A-3
Analyzer
module
Sample gas
Zero gas
Span gas
Electronics
module
System bus
Continued on next page