10-20 Chapter 10: Status Messages, Troubleshooting 42/24-10 EN Rev. 9
Caldos25, Caldos27, Magnos206, Magnos27 Problems
Temperature Problem
Faulty temperature
sensor or heater
connections
Check the connecting lines and plugs.
Check the line seating in the insulated jackets.
Defective thermal link Check thermal link continuity and replace if necessary (see page 9-3 for
instructions).
Leaking thermostat
or open purge gas
connections
Check the seal integrity between the thermostat chamber and the mounting
flange; tighten mounting bolts or replace O rings as needed.
Seal the analyzer purge gas inlet and outlet with sealing connectors.
Check the purge gas flow (maximum operating level of 20 l/h; positive pressure
p
e
= 2 to 4 hPa) and reduce as needed.
Unstable Readings
Gas path leakage Check the integrity of the analyzer module gas paths and (if applicable) of the
lines to the gas module (see page 9-2 for instructions).