42/24-10 EN Rev. 9 Chapter 9: Inspection and Maintenance 9-27
Changing Analog Output Current Range
Method The current range of the individual analog outputs can be modified by initializing
the applicable Analog output function blocks. Technical Bulletin “AO2000 Func-
tion Blocks – Descriptions and Configuration” (Publication No. 30/24-200 EN)
contains a detailed description of the function block.
Menu Path MENU
Configure Function blocks Outputs
Analog output
Figure 9-14
Analog Output
Function Block
Parameters
Changing Current
Range
The current range is changed using the User range parameter.
Selection The current range choices are 0-20 mA, 2-20 mA and 4-20 mA.
The output signal cannot be lower than 0 mA.