A-2-4 Appendix 2: Analyzer Module Operating Specifications 42/24-10 EN Rev. 9
Magnos206 Operating Specifications
Linearity deviation ≤ 0.5 % of span
Repeatability ≤ 50 ppm O
2
(time base for gas exchange ≥ 5 minutes)
Zero drift ≤ 3 % of span of the smallest measurement range (per order) per week, minimum
300 ppm O
2
per week; following prolonged transport and storage time the drift
can be higher during the first weeks of operation.
Sensitivity drift ≤ 0.1 % Vol.-% O
2
per week or ≤ 1 % of measured value per week (not cumulative),
the smaller value applies; ≤ 0.25 % of measured value per year
Output fluctuation (2 σ) ≤ 25 ppm O
2
at electronic T90 time (static/dynamic) = 3 /0 sec.
Detection limit (4 σ) ≤ 50 ppm O
2
at electronic T90 time (static/dynamic) = 3 /0 sec.
Flow influence ≤ 0.1 Vol.-% O
2
in permissible range
Associated gas
influence
The effect of associated gases with a concentration of 100 Vol.-% as a shift of the
zero-point (ΔZero) in Vol.-% O
2
can be estimated using the guidelines in the
following table:
Associated Gas Concentr. 100 Vol.-% ΔZero Associated Gas Concentr. 100 Vol.-% ΔZero
Hydrogen H
2
+0.28 Carbon dioxide CO
2
–0.32
Hydrogen sulfide H
2
S –0.45 Carbon oxysulfide COS –0.90
Argon Ar –0.26 Ethane C
2
H
6
–0.46
Helium He +0.30 Ethylene C
2
H
4
–0.29
Neon Ne +0.13 Methane CH
4
–0.24
Nitrogen N
2
0.00 Propane C
3
H
8
–0.98
Nitrogen oxide NO +43.00 Propylene C
3
H
6
–0.55
Nitrogen dioxide NO
2
+28.00 Trichloroethylene C
2
HCl
3
–2.17
Nitrous oxide N
2
O –0.20 Vinylchloride CH
2
CHCl –0.75
Carbon monoxide CO –0.01 For further associated gases refer to EN 61207-3
Temperature
influence
• At zero-point: ≤ 0.02 Vol.-% O
2
per 10 °C
• On sensitivity: ≤ 0.1 % of measured value per 10 °C
Ambient temperature in permissible range, thermostat temperature = 64 °C
Air pressure influence • At zero-point: no influence effect
• On sensitivity without pressure correction: ≤ 1 % of measured value per 1 % air
pressure change
• On sensitivity with pressure correction by means of internal pressure sensor
(option): ≤ 0.1 % of measured value per 1 % pressure change; for highly
suppressed measurement ranges ≤ 0.01 % of measured value per 1 % pressure
change or ≤ 0.002 Vol.-% O
2
per 1 % pressure change, whichever is greater.
Pressure sensor working range: p
abs
= 600 to 1250 hPa
Power supply
influence
24 VDC ± 5 %: ≤ 0.4 % of span
Position influence Zero-point shift ≤ 0.05 Vol.-% O
2
per 1° deviation from horizontal orientation.
Position has no effect on the hard-mounted unit.
T
90
time T
90
≤ 3.5 to 10 sec. at sample gas flow = 90 l/h and electronic T90 time (static/
dynamic) = 3/ 0 sec., gas switching from N
2
to air (applies to a gas analyzer with
one analyzer module)