1-2 Chapter 1: Preparing the Installation 42/24-10 EN Rev. 9
Installation Location Requirements, continued
Installation Location
Altitude
The maximum installation location altitude is 2000 m.
Vibrations If the gas analyzer is installed in a cabinet the maximum acceleration amplitude is
0.01 ms
–2
in a frequency range of 0.1 to 200 Hz.
If the gas analyzer is not installed in a cabinet the following data for the individual
analyzer modules apply.
Analyzer Module Vibrations
Caldos25 max. ±0.04 mm at 5 to 30 Hz
Caldos27 max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
Limas11 max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
Magnos206 max. ±0.04 mm at 5 to 20 Hz
Magnos27 max. ±0.04 mm at 5 to 60 Hz
Uras26 max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz,
slight transient effect on sample value in the region
of the beam modulation frequency
Note: Compliance with the metrological data can only be assured if data on
vibration amplitude and frequency range at the installation site are available and
suitable means are employed to decouple the analyzer.