42/24-10 EN Rev. 9 Chapter 10: Status Messages, Troubleshooting 10-13
Status Messages, continued
No. Text O E M F Reaction /Remark
360 Filter wheel 1 cannot be
initialized.
x x Notify service.
361 Filter wheel 2 cannot be
initialized.
x x Notify service.
362 The calibration filter wheel
cannot be initialized.
x x Notify service.
363 The Limas analyzer board
cannot be initialized.
x x Notify service.
364 A new lamp is installed and
the amplifiers settings are
optimized.
For information
365 The LIMAS amplifiers settings
are optimized.
For information
LS25
366 The LS25 analyzer has an
error.
x x
367 The LS25 analyzer has a
maintenance request.
x x
368 The LS25 analyzer starts
measurement.
x x
369 LS25 analyzer detector error # x x
370 The beam transmission
exceeds the permissible
range.
x Low transmission
371 The temperature input signal
exceeds the permissible
range.
x T out of range
372 The pressure input signal
exceeds the permissible
range.
x P out of range
373 The flow sensor signal
exceeds the permissible
range.
x
374 The detector signal is too low
for measurement.
x x Laser line-up error
375 The input current loop
(4-20mA) has an error.
x x T-read error, P-read error
376 This LS25 module is under
maintenance.
x x Communication with service software via RS232. No
measured values are transferred.
Uras
378 Chopper wheel jammed. x x Notify service.
379 Chopper wheel speed not OK. x x Notify service.
380 IR radiator or electronics
faulty.
x x Notify service.
381 High voltage at preamplifier
faulty.
x x Notify service.
Continued on next page