Error messages
140 FEX300, FEX500 OI/FEX300/FEX500-EN
Error no. /
Range
Text on the LCD display Cause Remedy
C175.013
Config.
Totalizer Reset
Check digital in
terminals 81,82
The function of the digital input (DI) is
set to "external totalizer reset" and the
digital input (DI) is set to high signal
(+24 V DC).
Set the digital input (DI) to low signal
(0 V DC).
C174.002
Config.
Pulse Simulation
selected on DO1
Switch off
Simulation Mode
Simulation mode is activated. In the "Process Alarm" menu,
deactivate simulation mode.
C172.004
Config.
Pulse Simulation
selected on DO2
Switch off
Simulation Mode
Simulation mode is activated. In the "Process Alarm" menu,
deactivate simulation mode.
C168.001
Config.
Logic Simulation
selected on DO1
Switch off
Simulation Mode
Simulation mode is activated. In the "Process Alarm" menu,
deactivate simulation mode.
C164.003
Config.
Logic Simulation
selected on DO2
Switch off
Simulation Mode
Simulation mode is activated. In the "Process Alarm" menu,
deactivate simulation mode.
C158.039
Config.
Simulation of
HART frequency
Switch off
Simulation Mode
Simulation mode is activated. In the "Process Alarm" menu,
deactivate simulation mode.
C154.018
Config.
Simulation
Digital In
Switch off
Simulation Mode
Simulation mode is activated. In the "Process Alarm" menu,
deactivate simulation mode.