Parameterization
OI/FEX300/FEX500-EN FEX300, FEX500 97
7.4.3 Menu: Config. Device
Menu / Parameter Value range Description
Device Setup
Access Control
Select the "Access Control" submenu.
Sensor
Select the "Sensor" submenu.
Transmitter
Select the "Transmitter" submenu.
Device Setup / Access Control
Standard Password Alphanumeric
Enter / change the password for the "Standard"
access level.
Advanced Password Alphanumeric
Enter / change the password for the "Advanced"
access level.
Read Only Switch Display only (ON / OFF)
Display of switch position of BR902 (hardware write
protection).
For further information read and observe chapter 7.2.4
“Hardware write protection”.
Custody Switch Display only (ON / OFF)
Display of the switch position of the calibration circuit
breaker
(must be activated for calibrated devices).
Device Setup / Sensor
Q
max
DN Display only
The displayed value is the flow at a velocity of 10 m/s.
The value is determined automatically via the selected
flowmeter size.
Qmax
Min. flow range: 0 ... 0.2 m/s
(0 ... 0.2 x Q
max
DN)
Max. flow range: 0 ... 20 m/s
(0 ...2 x Q
max
DN)
Select the flow range end value (flow range 1) for
forward and reverse flow.
Default setting: 1 x Q
max
DN.
Qmax 2 See Qmax
Select the flow range end value (flow range 2) for
forward and reverse flow.
Default setting: 1 x Q
max
DN, flow range 2 is
deactivated.
Important (Notice)
The switchover between the two measuring
ranges is done via the digital input or via the
menu "Device Setup / Sensor / Range Mode"
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.