Parameterization
98 FEX300, FEX500 OI/FEX300/FEX500-EN
Menu / Parameter Value range Description
Device Setup / Sensor (continued)
Range Mode MB Qmax activated
MB Qmax 2 activated
Manual switchover between flow ranges Qmax and
Qmax 2.
Sensor Location TAG
Alphanumeric, max.
20 characters
Enter the TAG number of the flowmeter sensor
(shown in the upper left of the process display).
Sensor TAG
Alphanumeric, max.
20 characters
Enter the TAG number of the sensor.
Device Setup / Transmitter
Units
Select the "Units" submenu.
Damping 0,02 ... 60 s Set the damping (the value relates to 1 T (Tau).
The value relates to a step flowrate change.
It affects the instantaneous value in the display and at
the current output.
Default setting: 1 second
Density 0,01 ... 5,0 g/cm³
If the flow is measured and indicated in the units g/s,
g/min, g/h, kg/s, kg/min, kg/h, kg/d, t/min, t/h, t/d, lb/s,
lb/min, lb/h and lb/d, a fixed density must be taken into
account for the calculations.
To convert the flowrate to mass flow units, a density
value from 0.01 to 5.0 g/cm³ can be entered.
Low Flow Cut Off
Select the "Low Flow Cut Off" submenu.
TX Location TAG
Alphanumeric, max.
20 characters
Enter the TAG number for the transmitter.
TX TAG
Alphanumeric, max.
20 characters
Enter the TAG number for the transmitter.
Operating Mode
Select the "Operating Mode" submenu.
System Zero
Select the "System Zero" submenu.
Noise Reduction Off
Mean Filter
Notch Filter
Lowpass V=Auto
Lowpass V=1
Activates noise reduction in case of unstable flow
signal. Activating noise reduction increases the
response time.
Factory setting: Off
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.