Electrical connections
OI/FEX300/FEX500-EN FEX300, FEX500 39
5 Electrical connections
5.1 Preparing and routing the signal and magnet coil cable
5.1.1 Cable with part number D173D027U01
Prepare both cable ends as shown.
Important (Notice)
Use wire end sleeves.
Wire end sleeves 0.75 mm
2
(AWG 19), for shielding (1S, 2S)
Wire end sleeves 0.5 mm
2
(AWG 20), for all other wires
The shields may not touch (signal short circuit).
Fig. 32: Flowmeter sensor side, dimensions in mm (inch)
Fig. 33: Transmitter side, dimensions in mm (inch)
L
1
maximum stripped length = 105 (4.10)
1 Measurement potential 3, green L2 = 70 (2.76)
2 Signal line E1, violet L2 = 60 (2.36)
3 Shield 1S L2 = 60 (2.36)
4 Shield 2S L2 = 60 (2.36)
5 Signal line, E2, blue L2 = 60 (2.36)
6 Data line, D2, yellow L2 = 70 (2.76)
7 Data line, D1, orange L2 = 70 (2.76)
8 Magnet coil, M2, red L2 = 90 (3.54)
9 Magnet coil, M1, brown L2 = 90 (3.54)
10Groundwire, steel
11SE clamp