Parameterization
90 FEX300, FEX500 OI/FEX300/FEX500-EN
7.4 Description of parameters
7.4.1 Menu: Easy Setup
Menu / Parameter Value range Description
Easy Setup "Easy Setup" Menu
Language
Deutsch, English, Français,
Español, Italiano, Dansk,
Svenska, Polski, Russki,
Zhongweng, Turkce
Select the menu language.
Q (Flowrate) Unit
l/s; l/min; l/h; ml/s; ml/min;
m3/s; m3/min; m3/h; m3/d;
Ml/d; ft3/s; ft3/min; ft3/h; ft3/d;
ugal/s; ugal/min; ugal/h;
ugal/d; Mugal/d; igal/s;
igal/min; igal/h; igal/d; bls/s;
bls/min; bls/h; bls/d; hl/h; g/s;
g/min; g/h; kg/s; kg/min; kg/h;
kg/d; t/min; t/h; t/d; lb/s;
lb/min; lb/h; lb/d; custom/s
Select the unit for the flow indicator.
Default setting: l/min
Qmax
Min. flow range: 0 ... 0.2 m/s
(0 ... 0.02 x Q
max
DN)
Max. flow range: 0 ... 20 m/s
(0 ...2 x Q
max
DN)
Select the flow range for forward and reverse flow.
Default setting: 1 x Q
max
DN.
Totalizer/Pulse Unit
m3; l; ml; ft3; hl; g; kg; t; lb;
igal; ugal; bls; Ml; Mugal;
custom
Select the unit for the flowmeters.
Default setting: l
Operation
Pulse Mode, Fullscale
Frequency
Select the operating mode for the digital output.
There are two operating modes available:
• „Pulse Mode“: In pulse mode, pulses per unit are
output (e.g., 1 pulse per m
3
).
• „Fullscale Frequency“: In frequency mode, a
frequency proportional to the flowrate is output.
The maximum frequency corresponding to the flow
range end value is configurable (max. 5.25 kHz).
Default setting: „Pulse Mode“
Pulses per Unit -
Display of the pulses per unit output by the digital
output. The max. possible number of pulses is 5250
per second.
Fullscale Frequency 0 ... 5250 Hz
Set the frequency for the flow range end value in
Fullscale Frequency operating mode.
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.