Parameterization
OI/FEX300/FEX500-EN FEX300, FEX500 93
7.4.2 Menu: Device Info
Important (Notice)
This menu is used only to display the device parameters. The parameters are displayed
independently of the configured access level, but cannot be changed.
Menu / Parameter Value range Description
Device Info
Sensor
Select the "Sensor" submenu.
Acquisition
Select the "Acquisition" submenu.
Analog Range
Select the "Analog Range" submenu.
Transmitter
Select the "Transmitter" submenu.
Device Info / Sensor
Sensor Type -
Type of flowmeter sensor (ProcessMaster 300 / 500,
HygienicMaster 300 / 500).
Important (Notice)
When commissioning the device, make sure that
the transmitter is assigned to the sensor
correctly. It is not possible to operate a flowmeter
sensor of the 300 series with a transmitter of the
500 series.
Sensor Model - Indication of model number (e.g., FEP315)
Sensor Size - Size of sensor.
QmaxDN
-
This value is the maximum flow at a velocity of 10
m/s.
The value is set automatically via the selected
flowmeter size.
Qmax - Set flow range end value for flow range 1.
Factory setting: Flow range 1 activated.
Qmax 2 - Set flow range end value for flow range 2.
Factory setting: Flow range 2 deactivated.
Important (Notice)
The switchover between the two measuring
ranges is done via the digital input or via the
menu "Config. Device / Sensor / 2 flow ranges“
Span Ss - Calibration value for the sensor (span)
Zero Sz - Calibration value for the sensor (zero)
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.