Parameterization
92 FEX300, FEX500 OI/FEX300/FEX500-EN
Menu / Parameter Value range Description
Easy Setup (continued) "Easy Setup" Menu
Cable length
0.01 ... 200 m
Enter the signal cable length between the transmitter
and the flowmeter sensor.
For devices with a compact design
(FEP311, FEH311, FEP315, FEH315) 0.01 m must be
entered.
Important (Notice)
The entry is required for FEP500, FEH500 if the
diagnostic functions are to be used.
When using the ScanMaster verification software
you also have to enter the signal cable length.
Easy Setup / System Zero
Submenu "System Zero"
Manual
Starts the manual zero adjustment.
Automatic
Starts the automatic zero adjustment.
Important (Notice)
Prior to starting the zero adjustment, make sure
that:
• There is no flow through the flowmeter sensor
(close all valves, shut-off devices, etc.)
• The flowmeter sensor is completely filled with
the fluid to be measured.
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.